Accuracy and precision of 3D cephalometric landmarks from biorthogonal plain-film x rays

Autor: Omar Macaraeg, B. H. Broadbent, Krishna Subramanyan, Mark G. Hans, Martin Palomo, Alexander Moullas, David Dean
Rok vydání: 1998
Předmět:
Zdroj: SPIE Proceedings.
ISSN: 0277-786X
Popis: Three dimensional (3D) plain film radiographic cephalometric analysis of boney skull landmarks may be used for patient diagnosis, treatment planning, prosthetic design, intra- operatively, and outcome assessment. To test the accuracy and reliability of 50 cephalometric landmarks, three dry human skulls, with and without metallic markers affixed to the landmarks, were digitized in our 3dCEPH software by 4 operators. The average inter-operator variability about mean landmark position, across all operators, for all 3 skull image pairs, was 3.33 mm. Ten landmarks exhibiting least variability were 1.15 mm average distance from the mean, including: B point 0.69 mm, Lower Incisal Edge 0.85 mm, and Anterior Nasal Spine 0.90 mm. The average rms error from the metallic fiducials for these 4 operators across all 50 landmarks, and 3 skulls was 5.03 mm. The 10 landmarks with the least variability exhibited 2.01 mm average distance from the fiducial, including: B point 1.69 mm, upper incisal edge 1.71 mm, lower incisal edge 1.78 mm. Additional studies are needed to test the robusticity of the hypothesis of homologous anatomy. Homology of landmarks is important to cephalometric comparisons between image pairs representing patient and 'normative,' pre- and post-surgical alteration, and different ages of the same patient.© (1998) COPYRIGHT SPIE--The International Society for Optical Engineering. Downloading of the abstract is permitted for personal use only.