Dental Implant Macro-Design Features Can Impact the Dynamics of Osseointegration

Autor: Amol Chaudhari, Joke Duyck, Judith de Rycker, Katleen Vandamme, Ignace Naert, Marcio Vivan Cardoso, Bart Van Meerbeek
Rok vydání: 2013
Předmět:
Zdroj: Clinical Implant Dentistry and Related Research. 17:639-645
ISSN: 1523-0899
DOI: 10.1111/cid.12178
Popis: Purpose The purpose of this study was to compare the clinical performance of two dental implant types possessing a different macro-design in the in vivo pig model. Materials and methods Titanium Aadva(TM) implants (GC, Tokyo, Japan) were compared with OsseoSpeed(TM) implants (Astra, Molndal, Sweden), with the Aadva implant displaying significant larger inter-thread dimensions than the OsseoSpeed implant. Implants were installed in the parietal bone of 12 domestic pigs and left for healing for either 1 or 3 months. Implant osseointegration was evaluated by quantitative histology (bone volume relative to the tissue volume [BV/TV]; bone-to-implant contact [BIC]) for distinct implant regions (collar, body, total implant length) with specific implant thread features. The Wilcoxon-Mann-Whitney nonparametric test with α = 0.05 was performed. Results An inferior amount of bone enveloping the Aadva implant compared with the OsseoSpeed implant was observed, in particular at the implant body part with its considerable inter-thread gaps (p Conclusion Although the Aadva implant displayed a clinically acceptable level of osseointegration, the findings demonstrate that implant macro-design features can impact the dynamics of implant osseointegration. Consideration of specific implant macro-design features should be made relative to the biological and mechanical microenvironment.