The design and implementation of the beam diagnostics control system for HIMM
Autor: | Kang Xincai, Yan Yin, Wei Li, Wei-Nian Ma, T. C. Zhao, S.P. Li, Yongchun Feng, Z.L. Zhao, Y.G. Nie, Yonghao Chen, Peng Li, Z. Xu, Min Li, J.M. Dong, Ruishi Mao, Yanshan Wang, Youjin Yuan |
---|---|
Rok vydání: | 2019 |
Předmět: |
Physics
Nuclear and High Energy Physics Detector Motion control 01 natural sciences Ion source Synchrotron 030218 nuclear medicine & medical imaging law.invention 03 medical and health sciences 0302 clinical medicine Data acquisition Reliability (semiconductor) law Control system 0103 physical sciences Electronic engineering Physics::Accelerator Physics 010306 general physics Instrumentation Beam (structure) |
Zdroj: | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 919:27-35 |
ISSN: | 0168-9002 |
Popis: | The Heavy Ion Medical Machine (HIMM) is a compact synchrotron based facility built by the Institute of Modern Physics (IMP) in China. This facility delivers carbon ion beams with kinetic energies up to 400 MeV/u for clinical applications. The beam diagnostics system for HIMM is based on 18 different types of monitors and is used to measure all relevant beam parameters from the ion source to the irradiation therapy terminals which are critical for the commissioning of the accelerator as well as during the regular operation to ensure high up-time of the machine. The control system for all the monitors is designed, developed and implemented globally based on a 3-tier architecture with the aim to achieve a clear separation between front-end devices and their controllers. This paper presents the architecture, the design principle and the implementation of this control system. The beam parameters measured based on the detectors ′ motion control and data acquisition control system are analyzed in detail. Furthermore, this beam diagnostics control system (BDCS) will be transplanted to the next heavy ion medical machine in Lanzhou city on account of the successful application in Wuwei city, the precise beam parameters measurement, and long term operation with high stability and reliability. |
Databáze: | OpenAIRE |
Externí odkaz: |