Popis: |
Plastic scintillators appear to have potential utility in three research areas related to nuclear medicine: (1) high count rate applications in general, (2) positron camera applications, and (3) positron source localization through measurement of relative arrival times of annihilation quanta at two co-linear detectors. These three areas of applicability depend on improvement in three specific areas of plastic scintillator technology: (a) development of plastics with very fast decay times, (b) development of plastics with greatly improved high energy photon detection efficiencies (high-Z loaded plastics), and (c) improvement of fast timing system capabilities. The three preceding areas of improvement are discussed. |