Imaging the Probe Skirt in the Environmental SEM
Autor: | I. C. Bache, P. Smith, B. L. Thiel |
---|---|
Rok vydání: | 2000 |
Předmět: | |
Zdroj: | Microscopy and Microanalysis. 6:794-795 |
ISSN: | 1435-8115 1431-9276 |
DOI: | 10.1017/s1431927600036461 |
Popis: | In low vacuum scanning electron microscopes, the primary beam is partially scattered by the gas present in the specimen chamber. The development of these microscopes, in particular the so-called ‘Environmental’ SEM, was initiated when it was realized that this scattering does not necessarily compromise the imaging capabilities of the instrument. Indeed, some modern commercial instruments are capable of better than 2 nanometer resolution at gas pressures of several torr. The accepted explanation for this is as follows: The mean-free-path of the high energy primary electrons is several millimeters in one torr of water vapour (for example). Because the actual pathlength of electrons travelling through the gas is only a few millimeters, most of them do not scatter at all. Those that do scatter are supposedly distributed over a relatively large area. Thus, the probe features a high-intensity central region surrounded by a slowly decaying low-intensity skirt. High resolution imaging is possible because the signal-to-(skirt)background ratio is high. |