Autor: |
M Simon, Cs Trajber, S Bohátka, I Futó |
Rok vydání: |
1993 |
Předmět: |
|
Zdroj: |
Vacuum. 44:653-656 |
ISSN: |
0042-207X |
DOI: |
10.1016/0042-207x(93)90117-s |
Popis: |
An improvement of a new computational method for mass independent optimization of quadrupole pre-filters is presented. Computer simulation of mass peaks and calculated ion-beam cross-sections support the predictions of the optimization procedure. Transmissions with respect to those of a single analyser are uniformly improved at pre-filter lengths corresponding to the minima of the suggested optimizing function at a fixed ion velocity. Measurements for light ions with the experimental setup constructed according to the proposed method are also in good agreement with the numerical results. |