Detection of Amyloid Beta Fibril Growth by Liposome-Immobilized Micro-Cantilever With NiCr Thin-Film Strain Gauge
Autor: | Kaoru Yamashita, Z. Zhang, Toshio Akai, Masayuki Sohgawa, Keisuke Takada, Minoru Noda |
---|---|
Rok vydání: | 2015 |
Předmět: | |
Zdroj: | IEEE Sensors Journal. 15:7135-7141 |
ISSN: | 2379-9153 1530-437X |
DOI: | 10.1109/jsen.2015.2470108 |
Popis: | A micro-cantilever with thermally stable strain gauge of a NiCr thin film and a self-assembled monolayer was fabricated by surface micromachining. Liposomes, as biosensing molecules, were immobilized on its surface. To prevent evaporation of the water solvent, a reservoir made of polydimethylsiloxane around the micro-cantilever was fabricated and covered with a glass plate. The resistance of the strain gauge increases with time in the amyloid beta ( $\text{A}\beta $ ) solution, because the cantilever is bent downward caused by interaction between $\text{A}\beta $ and liposomes immobilized on the micro-cantilever surface. Moreover, resistance change depends on the interaction ability of $\text{A}\beta $ proteins with liposomes. It is therefore indicated that the fabricated static mode micro-cantilever sensor with immobilized liposomes can be used to detect $\text{A}\beta $ proteins. |
Databáze: | OpenAIRE |