Push and Fluff technique for optimization of clot integration with stent-retriever: An in vitro model

Autor: Agostinho C. Pinheiro, Raul G. Nogueira, Ryan M. Grandfield, Shao-Pow Lin, Aniel Q. Majjhoo, Amin Nima Aghaebrahim, Michael G. Abraham, Paul Mazaris, Justin A. Singer, Alhamza R. Al-Bayati, Leonard H. Verhey, Eugene Lin, Diogo C. Haussen
Rok vydání: 2023
Předmět:
Zdroj: Interventional Neuroradiology. :159101992311753
ISSN: 2385-2011
1591-0199
DOI: 10.1177/15910199231175348
Popis: Background For stent–retriever (SR) thrombectomy, technical developments such as the Push and Fluff technique (PFT) appear to have a significant impact on procedural success. This study aimed to (1) quantify the enhancement in clot traction when using PFT as compared to the standard unsheathing technique (SUT) and (2) to evaluate the performance of PFT in new versus established users of the technique. Methods Operators were divided between established PFT and SUT users. Each experiment was labeled according to the SR size, utilized technique, and operator experience. A three-dimensional-printed chamber with a clot simulant was used. After each retriever deployment, the SR wire was connected to a force gauge. Tension was applied by pulling the gauge until clot disengagement. The maximal force was recorded. Results A total of 167 experiments were performed. The median overall force to disengage the clot was 1.11 pounds for PFT and 0.70 pounds for SUT (an overall 59.1% increment with PFT; p Conclusion PFT led to reproduceable improvement in clot engagement with an average ∼60% increase in clot traction in this model and was found not to have a significant learning curve.