Echocardiographic Guidance of Percutaneous Patent Foramen Ovale Closure: Head-to-Head Comparison of Transesophageal versus Rotational Intracardiac Echocardiography
Autor: | Cesare Amico, Francesco Loperfido, Michele Pacilli, Carlo Vigna, Nicola Marchese, Mario Fanelli, Raffaele Fanelli, Mario Zanchetta, Massimo Chessa, Vincenzo Inchingolo |
---|---|
Rok vydání: | 2012 |
Předmět: |
medicine.medical_specialty
Aorta Percutaneous business.industry Right-to-left shunt medicine.disease Intracardiac injection medicine.anatomical_structure Embolism Internal medicine medicine.artery medicine Patent foramen ovale Cardiology Radiology Nuclear Medicine and imaging Fossa ovalis Cardiology and Cardiovascular Medicine business human activities Shunt (electrical) |
Zdroj: | Echocardiography. 29:1103-1110 |
ISSN: | 0742-2822 |
DOI: | 10.1111/j.1540-8175.2012.01750.x |
Popis: | Background: Transesophageal (TEE) and intracardiac (ICE) echocardiography are commonly used to guide percutaneous patent foramen ovale (PFO) closure. The study aim was to perform a head-to-head comparison between TEE and rotational ICE echocardiography in the measurement of the fossa ovalis and device selection. Methods: In 45 patients with cryptogenic stroke or peripheral embolism and PFO with large right-to-left shunt, fossa ovalis dimensions were assessed preoperatively by TEE and intraoperatively by rotational ICE. The Amplatzer devices, deployed on the basis of ICE, were compared with those that would have been selected by TEE. Results: A good correlation between TEE and rotational ICE was observed for both longitudinal and transverse fossa ovalis dimensions (TEE four-chamber vs. ICE four-chamber: r = 0.75; TEE bicaval vs. ICE four-chamber: r = 0.77; TEE aorta vs. ICE aorta: r = 0.59; P < 0.001 for all). However, no such correlation was found in 13 patients with atrial septal aneurysm (ASA) (TEE four-chamber vs. ICE four-chamber: r = 0.33; TEE bicaval vs. ICE four-chamber: r = 0.49; TEE aorta vs. ICE aorta: r = 0.05; P = NS for all). At Bland-Altman analysis, slight systematic differences with wide limits of agreement for each comparison were observed, particularly in patients with ASA, suggesting that the two imaging modalities cannot be used interchangeably. As regards device selection, a moderate agreement was found between TEE- and ICE-guided device size (72%, κ= 0.53, P < 0.001), except in patients with ASA (36%, κ= 0.02, P = NS). Conclusions: Our study suggests a significant disagreement between TEE and rotational ICE in measuring fossa ovalis and selecting the device for PFO closure, particularly in patients with ASA. |
Databáze: | OpenAIRE |
Externí odkaz: |