Localization of breast tumor using holographic interferometry

Autor: C. William Loughry, HyunDae Hong, Daniel B. Sheffer
Rok vydání: 1997
Předmět:
Zdroj: SPIE Proceedings.
ISSN: 0277-786X
DOI: 10.1117/12.274065
Popis: For the purpose of locating breast lesions, we have utilized holographic interferometry (HI) to generate a fringe contour distortion superficial to an indwelling tumor. HI has been successfully utilized to examine material defects or dynamic characteristics of laminated structure. Application of HI technique in medical imaging has been reported. HI requires appropriate stressing mechanism to produce fringe patterns of reasonable density and contrast. Based on our research, it was found that pressure stressing has the advantages of providing both adequate density and patterns of fringe lines and an easily controlled stressing capacity. In our setup, both breasts of a subject hang within a Lucite pressure chamber while the intramural pressure of the chamber is raised by reducing volume of air inside. Our preliminary results found distorted fringe lines over breast lesions. The accuracy of the developed system in localizing breast lesions was determined by correlating results to those of x- ray mammography.© (1997) COPYRIGHT SPIE--The International Society for Optical Engineering. Downloading of the abstract is permitted for personal use only.
Databáze: OpenAIRE