Design and performance of an evolved gas analysis ion attachment mass spectrometer

Autor: Yuki Kitahara, Masamichi Tsukagoshi, Toshihiro Fujii, Márta Juhász, Seiji Takahashi
Rok vydání: 2010
Předmět:
Zdroj: Rapid Communications in Mass Spectrometry. 24:2625-2630
ISSN: 0951-4198
DOI: 10.1002/rcm.4682
Popis: We designed a simple evolved gas analysis (EGA) system to act as a sampler between solid samples at atmospheric pressure and the high vacuum inside a mass spectrometer. The newly designed stainless steel system is simple, small and rugged and fulfills all the basic requirements for EGA. The temperature is programmable with 60°C/min as the maximum heating rate and the temperature range is up to 600°C. With this system coupled with lithium ion attachment mass spectrometry (IAMS), it is possible to study the temperature-programmed decomposition of a number of solid materials by detecting any chemical species on a real-time basis. For illustrative purposes, EGA-IAMS experiments of polyethylene polymers have been conducted. Copyright © 2010 John Wiley & Sons, Ltd.