High dispersion spectrometer for time resolved Doppler measurements of impurity lines emitted during ISTTOK tokamak discharges

Autor: C.A.F. Varandas, R. B. Gomes, J. A. C. Cabral, Elena A. Sokolova, S. D. A. Reyes Cortes
Rok vydání: 2003
Předmět:
Zdroj: Review of Scientific Instruments. 74:2071-2074
ISSN: 1089-7623
0034-6748
DOI: 10.1063/1.1537039
Popis: A high dispersion, multichannel, visible spectrometer has been developed for the ISTTOK tokamak, aiming at measuring the evolution of both the edge ion temperature and plasma poloidal velocity, by analysis of a CIII spectral line. It is based on a double monochromator configuration with two high ruling densities (2700 l/mm) and 1 m focal length concave diffraction gratings, a commercial double microchannel plate image intensifier, a fast (f/#=1) relay lens built with two high diameter condensers in a back-to-back configuration, and a 2048 element photodiode charge coupled device optical multichannel analyzer. Laboratory tests have shown that a high dispersion has been achieved (≈1.65 A/mm), while keeping good temporal resolution (5 ms). Typical values measured for ion temperature (from Doppler broadening) were Ti≅25 eV while for poloidal velocity (from Doppler shift) vθ≅6 km/s.