Autor: |
D. Wegrzynek, B. Holyńska, J. Ptasiński |
Rok vydání: |
1994 |
Předmět: |
|
Zdroj: |
Applied Radiation and Isotopes. 45:409-414 |
ISSN: |
0969-8043 |
DOI: |
10.1016/0969-8043(94)90103-1 |
Popis: |
A PC-based program called ETM has been written for routine Energy Dispersive x-Ray Fluorescence (EDXRF) analysis using the fundamental parameter method. The program can be applied for different sources of primary radiation. Samples to be analyzed may contain unknown matrix. Simple, non-iterative algorithm is used for correction of the absorption and enhancement effects. A special procedure is applied for calculating the corrected concentrations of Pb and As when both elements are present in a sample. The results of analysis of three standard reference materials for three sources of primary radiation are presented. The program has been coded in Turbo Pascal 6.0. |