A method for assessing and correcting coincidence summing effects for germanium detector efficiency calibrations

Autor: D. M. Montgomery, G. A. Montgomery
Rok vydání: 1995
Předmět:
Zdroj: Journal of Radioanalytical and Nuclear Chemistry Articles. 193:71-79
ISSN: 1588-2780
0236-5731
DOI: 10.1007/bf02041918
Popis: The addition of54Mn and65Zn to a nine-radionuclide standard (containing241Am.,109Cd.57Co,139Ce,203Hg.113Sn,137Cs,88Y, and60Co) provides the capability to determine the extent of coincidence summing for gamma rays from88Y and60Co. A method for correcting the efficiency points at 1332 keV (60Co) and 1836 keV (88Y) for coincidence summing is presented.