An isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS)-based candidate reference measurement procedure for the quantification of levetiracetam in human serum and plasma
Autor: | Anja Kobel, Tobias Schierscher, Neeraj Singh, Linda Salzmann, Franziska Liesch, Friederike Bauland, Andrea Geistanger, Lorenz Risch, Christian Geletneky, Christoph Seger, Judith Taibon |
---|---|
Rok vydání: | 2023 |
Předmět: | |
Zdroj: | Clinical Chemistry and Laboratory Medicine (CCLM). |
ISSN: | 1437-4331 1434-6621 |
DOI: | 10.1515/cclm-2022-1038 |
Popis: | Objectives To develop an isotope dilution-liquid chromatography-tandem mass spectrometry (LC-MS/MS)-based candidate reference measurement procedure (RMP) for levetiracetam quantification in human serum and plasma. Methods Quantitative nuclear magnetic resonance spectroscopy (qNMR) was used to characterize the RMP material to ensure traceability to SI units. To quantify levetiracetam, an LC-MS/MS method was optimized using a C8 column for chromatographic separation following protein-precipitation-based sample preparation. Spiked matrix samples of serum and plasma were used to test selectivity and specificity. Matrix effects were determined by performing a post-column infusion experiment and comparing standard line slopes. Precision and accuracy were evaluated over 5 days. Measurement uncertainty was evaluated according to the Guide to the Expression of Uncertainty in Measurement (GUM). Results The RMP was proven to be highly selective and specific with no evidence of a matrix effect, allowing for quantification of levetiracetam within the range of 1.53–90.0 μg/mL. Intermediate precision was k=2). Conclusions We present a novel LC-MS/MS)-based candidate RMP for levetiracetam in human serum and plasma. Its expanded measurement uncertainty of ≤4.0% meets the clinical needs in levetiracetam monitoring. Utilizing qNMR to characterize levetiracetam reference materials allowed metrological traceability to SI units. |
Databáze: | OpenAIRE |
Externí odkaz: |