Initial Experience With the Pediatric Impella Device: A Feasibility Study in a Porcine Model
Autor: | Joseph M. Forbess, Kristine J. Guleserian, Melissa K. Webb, Vasiliki V Dimas, Matthew Riegel, Jian Wang, Thomas M. Zellers |
---|---|
Rok vydání: | 2016 |
Předmět: |
Body surface area
medicine.medical_specialty business.industry medicine.medical_treatment Ultrasound General Medicine 030204 cardiovascular system & hematology 03 medical and health sciences Catheter 0302 clinical medicine medicine.anatomical_structure Right Common Carotid Artery 030225 pediatrics Ventricular assist device Mitral valve Internal medicine Circulatory system medicine Cardiology Radiology Nuclear Medicine and imaging Cardiology and Cardiovascular Medicine business Impella |
Zdroj: | Catheterization and Cardiovascular Interventions. 88:582-589 |
ISSN: | 1522-1946 |
Popis: | Objective This study's objective was to evaluate insertion techniques and device fit of the pediatric version of the Impella ventricular assist device in swine which had similar sized carotids and left ventricles (LVs) as children weighing 10–20 kg. Background Options for minimally invasive circulatory support in children are limited. A modified device based on the current Impella 2.5 platform was created in an effort to provide minimally invasive circulatory support for children. Methods Animal studies (n = 10) were performed to determine technical feasibility of device implant via the right common carotid artery (RCCA) in swine with a carotid and LV size similar to children with a BSA |