Autor: Arnd B. Buchwald, Dirk Vollmann, Jens Stevens, Christina Unterberg
Rok vydání: 2001
Předmět:
Zdroj: Journal of Interventional Cardiac Electrophysiology. 5:477-485
ISSN: 1383-875X
DOI: 10.1023/a:1013262431932
Popis: Automatic atrial anti-tachy pacing (aATP) is a novel approach to treat paroxysmal/persistent atrial tachyarrhythmias in pacemaker patients. To evaluate the efficacy of aATP in terminating spontaneous atrial flutter/tachycardia episodes (AT), a dual-chamber stimulator with extensive diagnostic capabilities and programmable aATP therapies (AT500TM, Medtronic Inc.) was implanted in 30 patients with conventional pacing indications. During a mean follow-up time of 5.5 (1–12) months, aATP was delivered for 10494 AT. According to automatic device analysis, 8289 AT were treated with success (success-rate 79.0%). On 468 AT stored with the corresponding atrial EGM, an additional manual analysis was performed. The success-rate based on automatic analysis of these AT episodes (73.1%) was comparable to that found for all treated AT (79.0%), but manual EGM analysis revealed that only 209 of the 468 treated AT episodes (44.7%) were actually terminated by aATP. The aATP success-rate in the slower (cycle length 360–270ms) AT detection zone was significantly higher (73.8%, 62/84eps) than in the overlapping, faster (cycle length 270–220ms) AT zone (38.3%, 147/384eps, P