Mass test of AdvanSiD model ASD-NUV3S-P SiliconPMs for the Pixel Timing Counter of the MEG II experiment
Autor: | T. Cervi, Roberto Di Nardo, M. Rossella, M. Simonetta, S. Bariani, E. Romano, P. W. Cattaneo, Alessandro Menegolli, F. Vercellati, Michele Prata, C. Scagliotti, O. Barnaba |
---|---|
Rok vydání: | 2017 |
Předmět: |
Physics
Photomultiplier Pixel 010308 nuclear & particles physics business.industry Scintillator 01 natural sciences Particle detector Semiconductor detector Optics Time of arrival Silicon photomultiplier 0103 physical sciences 010306 general physics business Instrumentation Image resolution Mathematical Physics |
Zdroj: | Journal of Instrumentation. 12:C02024-C02024 |
ISSN: | 1748-0221 |
Popis: | The MEG II Timing Counter will measure the positron time of arrival with a resolution of 30 ps relying on two arrays of scintillator pixels read out by 6144 Silicon Photomultipliers (SiPMs) from AdvanSiD. They must be characterized, measuring their breakdown voltage, to assure that the gains of the SiPMs of each pixel are as uniform as possible, to maximize the pixel resolution. To do this an automatic test system that can measure sequentially the parameters of 32 devices has been developed. |