Magnetic resonance-guided near-infrared tomography of the breast
Autor: | John B. Weaver, Marvin M. Doyley, Ben Brooksby, Christine Kogel, Brian W. Pogue, Shudong Jiang, Steven P. Poplack, Keith D. Paulsen, Hamid Dehghani |
---|---|
Rok vydání: | 2004 |
Předmět: | |
Zdroj: | Review of Scientific Instruments. 75:5262-5270 |
ISSN: | 1089-7623 0034-6748 |
DOI: | 10.1063/1.1819634 |
Popis: | The design and implementation of a multispectral, frequency-domain near infrared tomography system is outlined, which operates in a MRI magnet for utilization of MR-guided image reconstruction of tissue optical properties. Using long silica optical fiber bundles, measurements of light transmission through up to 12 cm of female breast tissue can be acquired simultaneously with MRI scans. The NIR system utilizes six optical wavelengths from 660 to 850 nm using intensity modulated diode lasers nominally working at 100 MHz. Photomultiplier tube detector gain levels are electronically controlled on a time scale of 200 ms, thereby allowing rapid switching of the source to locations around the tissue. There are no moving parts in the detection channels and for each source position, 15 PMTs operating in parallel allow sensitivity down to 0.5 pW/cm2 at the tissue surface. Images of breast tissue optical absorption and reduced scattering coefficients are obtained using a Newton-type reconstruction algorithm to solv... |
Databáze: | OpenAIRE |
Externí odkaz: |