Turbidimetric Assay for Tetracyclines in Feeds Using a Microtiter Plate System

Autor: Marietta Sue Brady, Stanley E Katz
Rok vydání: 1991
Předmět:
Zdroj: Journal of AOAC INTERNATIONAL. 74:465-470
ISSN: 0004-5756
DOI: 10.1093/jaoac/74.3.465
Popis: The microtiter plate system for turbidimetric assay of chlortetracycline (CTC) and oxytetracycline (OTC) levels in feeds uses a 96 well microtiter plate, a multichannel pipette, and an ELISA reader to measure turbidity. Feeds are extracted for both tetracyclines using AOAC extraction systems. For CTC, the range of the standard curve is 0.001-0.005 μg CTC/ mL; for OTC, the range is 0.004-0.016 μg OTC/mL. Repeatability of CTC assays, as shown by the coefficient of variation (CV), ranged from 0.54 to 5.65% for same-day assays and from 2.01 to 9.39% for assays on different days. For OTC, CVs ranged from 2.69 to 10.01% for same-day assays and 3.24 to 9.08 % for different-day assays. Average recoveries for CTC were 108.7% for same-day assays and 106.8% for different-day assays; for OTC, average recoveries were 112.4% and 106.5% for same-day and different-day assays, respectively.
Databáze: OpenAIRE