Autor: |
M. W. Turner, Peter Richardson, Antony Robert Wilkes |
Rok vydání: |
2012 |
Předmět: |
|
Zdroj: |
Anaesthesia. 67:1119-1124 |
ISSN: |
0003-2409 |
DOI: |
10.1111/j.1365-2044.2012.07248.x |
Popis: |
Disconnection of an epidural catheter from its connector may result in patient harm and commonly requires resiting of the epidural. Clamp-connector designs such as the novel Portex EpiFuseTM potentially offer an improved safety profile over screw-cap designs such as the Tuohy-Borst, but comparative studies are limited. We therefore compared the tensile strength of EpiFuse and Tuohy-Borst connectors in a laboratory setting.We further sought to establish whether operator modification of the EpiFuse increased its vulnerability to disconnection. The median (IQR [range]) force to induce disconnection was 8.0 (4.1–12.8 [0.0–22.6]) N for Tuohy-Borst connectors and 16.4 (15.2–17.7 [5.7–18.9]) and 15.9 (15.0–16.9 [5.8–18.1]) N for standard and modified EpiFuse connectors, respectively (p < 0.0001). The Tuohy-Borst was also less likely to meet British Standard requirements (13 ⁄ 20 sets vs 19 ⁄ 20 and 20 ⁄ 20, p = 0.002). Modification of the EpiFuse did not affect lumen patency or connection strength. |