Frühlockerungen eines zementierten Hüftendoprothesenstiels aus Titan

Autor: T.K. Lichtinger, R. T. Müller, N. Schürmann
Rok vydání: 2000
Předmět:
Zdroj: Der Unfallchirurg. 103:956-960
ISSN: 1433-044X
0177-5537
DOI: 10.1007/s001130050652
Popis: Up to now there are no facts concerning the loosening rate of the cemented titanium stem (Trios). The central guiding canula and the longitudinal drilled stem shall ensure a symmetric cement mantle. In order to estimate the risk of aseptic loosening all Trios prostheses implanted in our clinic (n = 67; operations 1/95-11/96; max. follow-up: 3.1 years) were examined in a retrospective study and compared with a conventional femoral component (n = 141; Muller special; CoCr; operations 1977-1982). We used survivorship analysis to assess the results (Kaplan-Meier method; end point: revision). Compared to the conventional femoral stem the survival rate of the Trios prostheses was significantly worse (p = 0.0001). The probability of no revision after 3 years was 72% (conventional femoral component: 96%). Preoperative x-rays showed radiolucent lines surrounding the cement mantle. These findings as well as the loosening at the cement metal interface were confirmed intraoperatively. The explanted stems showed corresponding signs of abrasion on the surface. Our results prove that the application of cemented titanium stems might be problematic. With regard to the Trios prosthesis we recommend close routine check-ups to keep the chance of an early revision.
Databáze: OpenAIRE