Novel technique for high throughput measurement of active monooxygenase concentration

Autor: Anna Joëlle Ruff, Thomas Horn, Tim Welters, Jochen Büchs, Ulrich Schwaneberg
Rok vydání: 2016
Předmět:
Zdroj: Biotechnology and Bioengineering. 114:929-933
ISSN: 0006-3592
DOI: 10.1002/bit.26215
Popis: Absorbance measurements via transmitting light spectroscopy in microtiter plates are established for high throughput screening of biological systems. These measurements allow for the determination of important process parameters within a short time. However, absorbance determination via transmitted light measurements is not always feasible. As for carbon monoxide difference absorbance spectroscopy, used for concentration measurements of active P450 monooxygenases (P450s), security standards, and consistent gassing have to be addressed. In this study, a non-invasive online measuring principle for absorbance via scattered light is proposed. Based on optical fiber measurements, a decrease in scattered light signals at 450 nm wavelength of reflecting polymer particles is observed, and P450 concentrations are calculated. In this way, high throughput determination of P450 concentrations in a secure, gas-tight environment is realized. The designed method was successfully applied to concentration measurements and carbon monoxide (CO) saturation kinetics ranging from 0.3 to 5.0 μM P450 BM3 achieving a measurement accuracy of ±0.05 μM P450. Biotechnol. Bioeng. 2017;114: 929-933. © 2016 Wiley Periodicals, Inc.