Scalable Detector Design for a High-Resolution Fast-Neutron Radiography Panel

Autor: Christian Young, Chloe Browning, Ryan Thurber, Matthew Smalley, Michael Liesenfelt, Jason Hayward, Nicole McFarlane, Jeff Preston
Rok vydání: 2023
DOI: 10.21203/rs.3.rs-2688845/v1
Popis: Development of a scalable flat-panel neutron radiography device is needed to meet the nondestructive testing needs of a growing industrial market. Flood field images from a Cs-137 gamma source and a Cf-252 fission neutron source were generated using a detector composed of a 3-mm-thick sheet of EJ-200, and 3-mm-thick sheet of acrylic light spreader, an 8 × 8 array of SensL MICROFJ-60035-TSV SiPMs, and an IDEAS ROSSPAD readout module. The readout module can be tiled together with similar systems to create a panel of any size, and is only limited by the number of available Ethernet ports in a switching system. A collimated gamma line image demonstrated a 90% spatial resolution of approximately 0.47 line pairs per centimeter and a 10% spatial resolution of approximately 2.32 line pairs per centimeter. A neutron edge image demonstrated a 90% spatial resolution of approximately 0.70 line pairs per centimeter and a 10% spatial resolution of approximately 3.35 line pairs per centimeter. Both of these images show the ability to generate radiographs with sub-SiPM spatial resolution. Using these readout modules, a large-scale radiographic panel can be developed by tiling ROSSPAD modules together.
Databáze: OpenAIRE