INTRAOPERATIVE BREAKAGE OF A REUSED 25 GAUGE ILM PEELING FÓRCEPS
Autor: | Nishita Yadav, Madhurima Roy, Aniruddha Maiti, Sangeeta Roy |
---|---|
Rok vydání: | 2021 |
Předmět: |
Pars plana
medicine.medical_specialty genetic structures business.industry medicine.medical_treatment Internal limiting membrane Forceps Vitrectomy Fundus (eye) medicine.disease Ilm peeling eye diseases body regions Left eye medicine.anatomical_structure Ophthalmology medicine sense organs Epiretinal membrane business |
Zdroj: | International Journal of Retina. 4:178 |
ISSN: | 2614-8536 2614-8684 |
Popis: | A 59 years male presented with painless diminution of vision in right eye for 4 months. His vision was 20/100 in right eye and 20/30 in left eye. Fundus examination revealed epiretinal membrane in right eye with central macular thickness of 747 μm for which he underwent 25 guage pars plana vitrectomy. Intraoperatively after internal limiting membrane peeling, during removal of the pieces of the membrane from the forceps with the endoilluminator pipe, accidentally one blade from the tip of ILM peeling forceps broke. In this report, we describe first case of inadvertent intraoperative fracture of 25- gauge Grieshaber DSP End-Grasping ILM forceps in a case of ERM who underwent vitrectomy and removal of ERM with successful removal of fractured tip of ILM forceps. |
Databáze: | OpenAIRE |
Externí odkaz: |