Autor: |
Sachin Junnarkar, Veljko Radeka, Paul O'Connor, Sudeepti Southekal, Martin Purschke, Stephen L. Dewey, R. Fontaine, S. Krishnamoorthy, D. Schlyer, Roger Lecomte, John C. Neill, Paul Vaska, S.-J. Park, Wynne K. Schiffer, S. P. Stoll, C. Woody, A. Kandasamy, Jean-Francois Pratte, Dianne E. Lee, Sri Harsha Maramraju, A. Kriplani |
Rok vydání: |
2007 |
Předmět: |
|
Zdroj: |
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 571:14-17 |
ISSN: |
0168-9002 |
DOI: |
10.1016/j.nima.2006.10.019 |
Popis: |
The RatCAP is a small, head-mounted PET tomograph designed to image the brain of a conscious rat without the use of anesthesia. The detector is a complete, high-performance 3D tomograph consisting of a 3.8 cm inside-diameter ring containing 12 block detectors, each of which is comprised of a 4×8 array of 2.2×2.2×5 mm 3 LSO crystals readout with a matching APD array and custom ASIC, and has a 1.8 cm axial field of view. Construction of the first working prototype detector has been completed and its performance characteristics have been measured. The results show an intrinsic spatial resolution of 2.1 mm, a time resolution of ∼14 ns FWHM, and a sensitivity of 0.7% at an energy threshold of 150 keV. First preliminary images have been obtained using 18 F-FDG and 11 C-methamphetamine, which show comparable image quality to those obtained from a commercial MicroPET R4 scanner. Initial studies have also been carried out to study stress levels in rats wearing the RatCAP. |
Databáze: |
OpenAIRE |
Externí odkaz: |
|