Autor: |
Andrew Pargellis, Ben D. Gardner, John Granahan, David E. Burchfield, Phillip M. Erwin, Chris Matty, Grey Gentry, Souzan Thoresen |
Rok vydání: |
2013 |
Předmět: |
|
Zdroj: |
43rd International Conference on Environmental Systems. |
DOI: |
10.2514/6.2013-3527 |
Popis: |
The Major Constituent Analyzer (MCA) is a mass spectrometer based system that measures the major atmospheric constituents on the International Space Station. In 2011, two MCA ORU 02 analyzer assemblies experienced premature on-orbit failures. These failures were determined to be the result of off-nominal ion source filament performance. Recent product improvements to ORU 02 designed to improve the lifetime of the ion pump also constrained the allowable tuning criteria for the ion source filaments. This presentation describes the filament failures as well as the corrective actions implemented to preclude such failures in the future. |
Databáze: |
OpenAIRE |