Calibration of three capnographs for use with helium and oxygen gas mixtures*

Autor: J. A. S. Ball, RM Grounds
Rok vydání: 2003
Předmět:
Zdroj: Anaesthesia. 58:156-160
ISSN: 0003-2409
DOI: 10.1046/j.1365-2044.2003.02963.x
Popis: Capnography is considered essential in the management of mechanically-ventilated patients. Helium, as an adjunct to mechanical ventilation, is the subject of renewed interest and used increasingly. However, helium affects the performance of infrared capnometry. We constructed a simple device to generate variable mixtures of helium, oxygen and carbon dioxide within the normal physiological range, and tested the performance of two side-stream and one in-line capnographs. We found that addition of helium to the gas mixture caused all three capnographs to underestimate the concentration of carbon dioxide. The underestimation increased as the proportion of helium increased. The maximum underestimation (30%) occurred in a 79:21 helium/oxygen mixture.
Databáze: OpenAIRE