A Study of The Correlation of The Area Dose with Residual CT Contrast Media and MRI Contrast Media during The Use of General Imaging Automatic Exposure Control System

Autor: Chan-Woo Hong, Jin-Hun Park, Young-Deuk Seo, Jung-Min Lee
Rok vydání: 2016
Předmět:
Zdroj: Journal of the Korean Society of Radiology. 10:619-627
ISSN: 1976-0620
Popis: The purpose of this study is to investigate the effect of CT contrast agent and MRI contrast agent on the area dose in the body when using automatic exposure control system in general radiography. After making rectangular holes in the center of the abdominal thickness paraffin phantom, CT contrast agent and MRI contrast agent were respectively diluted with physiological saline solution for contrast medium dilution ratio of 10:0, 9:1, 8:2, 7:3, 6:4, 5:5, 4:6, 3:7, 2:8, 1:9, 0:10%. Each experiment was set to 78 kVp, 320 mA, which is the proper condition for KUB photography, and thereafter a total of 30 inspections were made for each dilution ratio using an automatic exposure control device, and the area dose corresponding to the dilution ratio of each contrast agent, Average comparison and correlation analysis were performed on the exposure index. As a result, the CT contrast agent and the MRI contrast agent appeared different in area dose according to the dilution ratio(p