A Study of The Correlation of The Area Dose with Residual CT Contrast Media and MRI Contrast Media during The Use of General Imaging Automatic Exposure Control System
Autor: | Chan-Woo Hong, Jin-Hun Park, Young-Deuk Seo, Jung-Min Lee |
---|---|
Rok vydání: | 2016 |
Předmět: |
030506 rehabilitation
business.industry MRI contrast agent Radiography media_common.quotation_subject Imaging phantom Correlation 03 medical and health sciences Contrast medium 0302 clinical medicine Correlation analysis Contrast (vision) Medicine 030211 gastroenterology & hepatology 0305 other medical science business Nuclear medicine media_common Automatic exposure control |
Zdroj: | Journal of the Korean Society of Radiology. 10:619-627 |
ISSN: | 1976-0620 |
Popis: | The purpose of this study is to investigate the effect of CT contrast agent and MRI contrast agent on the area dose in the body when using automatic exposure control system in general radiography. After making rectangular holes in the center of the abdominal thickness paraffin phantom, CT contrast agent and MRI contrast agent were respectively diluted with physiological saline solution for contrast medium dilution ratio of 10:0, 9:1, 8:2, 7:3, 6:4, 5:5, 4:6, 3:7, 2:8, 1:9, 0:10%. Each experiment was set to 78 kVp, 320 mA, which is the proper condition for KUB photography, and thereafter a total of 30 inspections were made for each dilution ratio using an automatic exposure control device, and the area dose corresponding to the dilution ratio of each contrast agent, Average comparison and correlation analysis were performed on the exposure index. As a result, the CT contrast agent and the MRI contrast agent appeared different in area dose according to the dilution ratio(p |
Databáze: | OpenAIRE |
Externí odkaz: |