Electrothermal Atomization Atomic Absorption Spectrometry with Platinum Tube Atomizer in Air
Autor: | Kiyohisa Ohta, Syn-ichi Itoh, Takayuki Mizuno |
---|---|
Rok vydání: | 1992 |
Předmět: |
Detection limit
Chemistry Biochemistry (medical) Clinical Biochemistry Analytical chemistry chemistry.chemical_element Alkali metal Biochemistry Analytical Chemistry law.invention Improved performance law In situ analysis Electrochemistry Trace analysis Tube (fluid conveyance) Atomic absorption spectroscopy Platinum Spectroscopy |
Zdroj: | Analytical Letters. 25:745-752 |
ISSN: | 1532-236X 0003-2719 |
DOI: | 10.1080/00032719208020031 |
Popis: | An electrothermal atomization method of atomic absorption spectrometry using a platinum tube atomizer in air is described. The atomizer consists of 0.025 mm-walled platinum tube and two supports. The remarkable merit of the platinum atomizer is the stability in air, when heated. This paper describes the improved performance for volatile elements like alkalimetals. Detection limits and characteristic mass are calculated for 7 elements and compared with literature values. The platinum atomizer can be used with a simple, effective and relatively low-cost atomic absorption spectrometer for routine in situ analysis. |
Databáze: | OpenAIRE |
Externí odkaz: |