A digitally assisted coaxial bridge—mainly consisted of a ratio transformer and digital voltage sources—was set up in a 4-terminal-pair (4-TP) configuration for the precision capacitance measurements. The 10: 1 capacitance ratio was measured in the range of 1 pF $\sim 1\ \mu\mathrm{F}$ mainly at 1 kHz using the 4-TP bridge. The measurement results show a degree of equivalence from a few parts in 108 to 105 when compared with those obtained using 2-terminal-pair/3-terminal bridges. Moreover, the measurement frequency of the 4-TP bridge was varied from 50 Hz to 20 kHz. The Allan deviation measurements imply that the statistical uncertainty could be achieved a level of 10−9 at 1 kHz. |