Trace metal atomic absorption spectrometric analysis utilizing sorbent extraction on polymeric-based supports and renewable reagents
Autor: | Encarnación Rodrı́guez Gonzalo, Herbert L. Lancaster, Graham D. Marshall, Jaromír Růžička, Gary D. Christian |
---|---|
Rok vydání: | 1994 |
Předmět: | |
Zdroj: | The Analyst. 119:1459-1465 |
ISSN: | 1364-5528 0003-2654 |
DOI: | 10.1039/an9941901459 |
Popis: | Various methods of analyte preconcentration for atomic absorption spectrometry are available for trace metals in aqueous samples. Among the drawbacks of many of these techniques are the generation of hazardous or toxic wastes, poor preconcentration efficiencies and cumbersome on-line implementation of the particular technique. Preconcentration of trace metals by sorbent extraction utilizing a flow injection method is described. Preconcentration of the analyte is accomplished by its extraction on a solid-phase hydrophobic support prepared by adsorption of a suitable chelating agent. Following extraction of the analyte, the metal–chelate complex is eluted from the solid support by changing the polarity of the carrier stream. The metal systems chosen for examination were copper and lead chelates of diethyldithiocarbamate (DDC) and quinolin-8-ol. Some results are presented for other chelates, namely 4-(2-pyridylazo)resorcinol, 1-(2-pyridylazo)-2-naphthol and dithizone. The sample throughput of the method was 20 h–1 utilizing 2 ml of sample per analysis for preconcentration. The detection limits using the quinolin-8-ol and DDC chelates are at low parts-per-billion levels, with enhancement factors between 50 and 100. |
Databáze: | OpenAIRE |