Monte Carlo simulation of pure tone audiometry reveals threshold-dependent differences in error
Autor: | Trevor T. Perry, Douglas S. Brungart |
---|---|
Rok vydání: | 2023 |
Předmět: | |
Zdroj: | The Journal of the Acoustical Society of America. 153:A47-A47 |
ISSN: | 1520-8524 0001-4966 |
Popis: | This study used a Monte Carlo simulation to evaluate the reliability of pure tone audiometry while incorporating information about the population distribution of thresholds. Results confirm that audiometric error (outcome threshold minus modeled threshold) is symmetric when simulated data are binned by modeled threshold. However, when data are binned by outcome threshold, the error depends on the relative distribution of modeled thresholds. When a realistic distribution is used, outcome thresholds of 0 dB HL show the lowest error rate. Bins with outcome thresholds 0 dB HL show a positive median error (outcome threshold > modeled threshold). These results suggest that a substantial proportion of individuals who obtain a very good threshold during audiometric testing (−5 or −10 dB HL) are individuals who actually have higher thresholds, which are being underestimated. These findings are relevant for hearing conservation programs, which monitor for upward shifts from baseline thresholds. [The views expressed in this abstract are those of the authors and do not reflect the official policy of the Department of Army/Navy/Air Force, Department of Defense, or U.S. Government.] |
Databáze: | OpenAIRE |