Sensitivity and Specificity of the BAX® for Screening/Listeria monocytogenes Assay: Internal Validation and Independent Laboratory Study
Autor: | W. M. Barbour, Curiale M, A. M. Hochberg, Gangar, Roering A, Peter M. Mrozinski |
---|---|
Rok vydání: | 2001 |
Předmět: | |
Zdroj: | Journal of AOAC INTERNATIONAL. 84:1087-1098 |
ISSN: | 1944-7922 1060-3271 |
DOI: | 10.1093/jaoac/84.4.1087 |
Popis: | In this study to certify the BAX® for Screening/Listeria monocytogenes assay (DuPont Qualicon, Wilmington, DE), an internal evaluation was conducted on 16 food types that were simultaneously analyzed with the BAX system (BAX), and the ISO method for the detection of L. monocytogenes (ISO). No statistically significant difference in performance between the BAX and ISO methods was observed. Inclusivity/exclusivity testing showed that the BAX system was able to detect 97 of 97 (100%) of L. monocytogenes strains tested. None of 56 other Listeria species or non-Listeria tested gave a reproducible positive BAX result. Ruggedness testing demonstrated that performance of the assay was not affected by reasonable variability in the operating parameters. BAX was then submitted for independent laboratory validation. In this phase, BAX was compared with standard culture methods for the detection of L. monocytogenes in chicken (USDA-FSIS), crab meat (BAM), and milk (AOAC). This study validated product claims of sensitivity and specificity >98% in accordance with AOAC Performance Tested MethodSM requirements. |
Databáze: | OpenAIRE |
Externí odkaz: |