Severe Hemolysis after Implantation of a Rashkind Occluder in a Large Ductus

Autor: Jonathan K.S. Ho, Gary Yk Mak, Lawrence Kk Tse, John E. Sanderson, Anthony P.C. Yim, Janet Fy Lee, J. M. Low
Rok vydání: 1997
Předmět:
Zdroj: Asian Cardiovascular and Thoracic Annals. 5:56-58
ISSN: 1816-5370
0218-4923
Popis: We report an unusual case of severe mechanical hemolysis after implantation of a 17-mm Rashkind patent ductus arteriosus occluder in a young adult female with a large ductus. The device failed with conservative management, including implantation of a second occluder, and surgery was required to close the ductus. This occluder may not be suitable for closure of a large patent ductus arteriosus.