Minimal vascular flows cause strong heat sink effects in hepatic radiofrequency ablation ex vivo

Autor: Kai S. Lehmann, Christian Rieder, Urte Zurbuchen, Christoph Holmer, Jörg P. Ritz, Andrea Schenk, Franz Poch, Ole Gemeinhardt, Andrea Stroux, Bernd Frericks, Martin E. Kreis
Rok vydání: 2016
Předmět:
Zdroj: Journal of Hepato-Biliary-Pancreatic Sciences. 23:508-516
ISSN: 1868-6974
DOI: 10.1002/jhbp.370
Popis: Background The present paper aims to assess the lower threshold of vascular flow rate on the heat sink effect in bipolar radiofrequency ablation (RFA) ex vivo. Methods Glass tubes (vessels) of 3.4 mm inner diameter were introduced in parallel to bipolar RFA applicators into porcine liver ex vivo. Vessels were perfused with flow rates of 0 to 1,500 ml/min. RFA (30 W power, 15 kJ energy input) was carried out at room temperature and 37°C. Heat sink effects were assessed in RFA cross sections by the decrease in ablation radius, area and by a high-resolution sector planimetry. Results Flow rates of 1 ml/min already caused a significant cooling effect (P ≤ 0.001). The heat sink effect reached a maximum at 10 ml/min (18.4 mm/s) and remained stable for flow rates up to 1,500 ml/min. Conclusions Minimal vascular flows of ≥1 ml/min cause a significant heat sink effect in hepatic RFA ex vivo. A lower limit for volumetric flow rate was not found. The maximum of the heat sink effect was reached at a flow rate of 10 ml/min and remained stable for flow rates up to 1,500 ml/min. Hepatic inflow occlusion should be considered in RFA close to hepatic vessels.