Autor: |
Michael Partridge, Nigel Guerra, Grace W. Muna, Holly Garner, Hala Sirhan, Bridget VerVaet |
Rok vydání: |
2014 |
Předmět: |
|
Zdroj: |
Electroanalysis. 26:2145-2151 |
ISSN: |
1040-0397 |
Popis: |
This paper demonstrates the development of an analytical method for detecting steroid hormones by coupling HPLC to electrochemical detection, using a nickel-modified glassy carbon electrode. The method was evaluated in terms of sensitivity, linear dynamic range, limit of detection, and response stability. The developed method exhibited good figures of merit for the steroid hormones studied with no evidence of electrode fouling. As an example, the limit of detection (S/N=3) for E3 was 0.10 µM and the response precision (n=5) was 0.6 %. The application of the method for the analysis of a real river water sample is demonstrated. |
Databáze: |
OpenAIRE |
Externí odkaz: |
|
Nepřihlášeným uživatelům se plný text nezobrazuje |
K zobrazení výsledku je třeba se přihlásit.
|