A Repetitional Pulsed X-Ray Diffractometer System
Autor: | Hikaru Terauchi, Toshio Kato, Kanji Kubota, Minoru Den, Toshihiro Kunimatsu, Norihiko Ninomiya, Toshiro Kojima, Yasusada Yamada, Kiyoshi Sakaue, Nobuyuki Nakamura, Satoshi Iida, Hiroshi Yoshimoto |
---|---|
Rok vydání: | 1983 |
Předmět: | |
Zdroj: | Japanese Journal of Applied Physics. 22:1444 |
ISSN: | 1347-4065 0021-4922 |
DOI: | 10.1143/jjap.22.1444 |
Popis: | A pulsed X-ray source and a diffractometer used with this source have been constructed. The pulsed X-ray tube can be driven at a peak voltage of 300 kV and a peak anode current of 5 kA to give X-ray pulses with a duration of about 40 ns at a repetition rate of up to 10 Hz. The high voltage is generated by a Marx generator coupled to a pulse forming line. One X-ray pulse contains about 3×106 Cu Kα photons. Some preliminary observations of the diffraction patterns from an LiF crystal have been made using a multichannel photodiode array as a detector. The use of this apparatus as a time-resolved X-ray diffractometer are discussed in this report. |