The accuracy of the ATLAS muon X-ray tomograph
Autor: | C. Boudineau, L Vertogradov, E. Sbrissa, R. Avramidou, C.W. Fabjan, E. Sedykh, Yury Smirnov, J. Berbiers, C Dechelette, S Grau, Edda Gschwendtner, Irene Vichou, H. Rieder, F. Rohrbach, S. Rangod, J. M. Maugain, I Sedykh, D Drakoulakos |
---|---|
Rok vydání: | 2003 |
Předmět: |
Physics
Nuclear and High Energy Physics Accuracy and precision Muon Large Hadron Collider Physics::Instrumentation and Detectors business.industry ATLAS experiment Detector Nuclear physics medicine.anatomical_structure Optics Atlas (anatomy) Scintillation counter medicine High Energy Physics::Experiment Tomography business Instrumentation |
Zdroj: | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 496:83-101 |
ISSN: | 0168-9002 |
DOI: | 10.1016/s0168-9002(02)01611-x |
Popis: | A gigantic detector, the ATLAS project, is under construction at CERN for particle physics research at the Large Hadron Collider which is to be ready by 2006. An X-ray tomograph has been developed, designed and constructed at CERN in order to control the mechanical quality of the ATLAS muon chambers. We reached a measurement accuracy of 2 μm systematic and 2 μm statistical uncertainties in the horizontal and vertical directions in the working area 220 cm (horizontal)×60 cm (vertical). Here we describe in detail the fundamental approach of the basic principle chosen to achieve such good accuracy. In order to crosscheck our precision, key results of measurements are presented. |