Neutron spectra around medical treatment facilities
Autor: | F. M. Wagner, Peter Kneschaurek, G. Schraube, E. Weitzenegger, H Schraube |
---|---|
Rok vydání: | 2002 |
Předmět: |
Bonner sphere
Physics Nuclear and High Energy Physics Spectrometer business.industry Astrophysics::High Energy Astrophysical Phenomena Nuclear Theory Linear particle accelerator Nuclear physics Neutron detection Research reactor Neutron Irradiation Radiation protection Nuclear Experiment business Instrumentation |
Zdroj: | Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment. 476:463-467 |
ISSN: | 0168-9002 |
DOI: | 10.1016/s0168-9002(01)01491-7 |
Popis: | The neutron field condition in and around the treatment room of the neutron therapy facility at the research reactor FRM in Garching-Munchen was studied for radiation protection purposes. Neutron spectrometry was conducted by using a 15-channel multisphere spectrometer. From the unfolded spectral shapes, averaged dose conversion coefficients were derived and correction factors for the reading of two neutron monitors obtained. These results were compared with further measurements done at a medical linear accelerator where the neutrons are produced by (γ, n) processes in the irradiation head. |