The effect of cooling rate on the performance of thermoluminescent dosimeter crystals (TLD-100) and LiF crystals
Autor: | E Lilley, M H Bradbury, B C E Nwosu |
---|---|
Rok vydání: | 1976 |
Předmět: | |
Zdroj: | Journal of Physics D: Applied Physics. 9:1009-1016 |
ISSN: | 1361-6463 0022-3727 |
DOI: | 10.1088/0022-3727/9/6/013 |
Popis: | A detailed study has been made of the effect of cooling rate on (a) the thermoluminescent sensitivity of TLD-100 and Harshaw pure LiF, and (b) the Mg++V associated pair concentration in TLD-100. The thermoluminescence of TLD-100 at the lower cooling rates appears to be strongly affected by precipitation of a second phase, whereas changes at higher cooling rates appear to be a consequence of Mg++V clustering. In particular there is evidence to show that peaks II and III are probably associated with the same trapping centre and that this is the Mg++V pair. |