Installation and measurement capacity of 3 × 592 GBq 241Am–Be neutron irradiation cell

Autor: I. Çelenk, S. Bulut, S. Artan
Rok vydání: 2013
Předmět:
Zdroj: Kerntechnik. 78:519-525
ISSN: 2195-8580
0932-3902
Popis: In this study, the installation and measurement capacity of the neutron irradiation system are investigated. First of all an irradiation geometry enabling optimum irradiation was designed for three 241Am–Be sources each of it having 592 GBq activity. Neutron irradiation system was installed after design and optimization of the system including the design of appropriate moderator and shielding were completed. Radiation safety standards of the Neutron Research Laboratory fulfilling the requirements of national regulation were achieved with unique configuration of the shielding materials. In this study the results of qualitative and quantitative detection limits obtained for Na, Al, Cl, K, Ti, V, Mn, Fe, Co, Cu, Zn, Ga, Ge, As, Se, Br, Rb, Sr, Y, Mo, Ru, Ag, Cd, In, Sb, I, Cs, Ba, La, Ce, Pr, Nd, Sm, Eu, Gd, Dy, Ho, Yb, Lu, Hf, W, Pt, Au, Th and U elements by using the neutron irradiation cell comprising 3 × 592 GBq 241Am–Be isotopic neutron source are presented and discussed.
Databáze: OpenAIRE