Application of the sum-peak method to activity standardizations of 152 Eu sources in LNMRI (BR)

Autor: C.J. da Silva, Mam Araujo, P.A.L. da Cruz, C.F. Arcanjo, Ricardo Tadeu Lopes, A. L. Ferreira Filho, José Ubiratan Delgado, R.W.R.V. Da Silva, R. Poledna
Rok vydání: 2018
Předmět:
Zdroj: Applied Radiation and Isotopes. 134:340-350
ISSN: 0969-8043
Popis: The sum-peak method, an absolute measurement technique that uses coincidence counting and γ-ray spectrometry, was applied to activity standardization of 152Eu sources. The decay branch used was that of 152Sm, as it is almost entirely electron capture and exhibits X-rays with a high probability of coincidence with the 121.8 keV gamma rays, as evidenced by the matrix technique used for complex decay schemes, a method for deriving counting rate equations describing coincidence summing of gamma and X-rays. The result was compared with the calibration performed by ionization chamber traceable to BIPM. The results presented uncertainty values of 0.50% (k = 1).
Databáze: OpenAIRE