Autor: |
L.A. Christian, M.D. Early, B. Moore |
Rok vydání: |
2008 |
Předmět: |
|
Zdroj: |
2008 Conference on Precision Electromagnetic Measurements Digest. |
DOI: |
10.1109/cpem.2008.4574970 |
Popis: |
We investigate the causes of the observed normal mode offsets in precision DMM's. With a circuit simulation tool to model the front-end amplifier we have been able to qualitatively reproduce the observed offset in a commonly used 8.5 digit precision DMM. In general the cause of the offset is related to the symmetry of the front-end architecture. However it seems there are several mechanisms which contribute to the effect, and it is these that we are presently seeking to elucidate. |
Databáze: |
OpenAIRE |
Externí odkaz: |
|