Assessment of a press-fit proximal femoral modular reconstruction implant (PFMR®) at 14.5 years. A 48-case series with a disturbing rate of implant fracture
Autor: | Daniel Molé, Adrien Jacquot, François Sirveaux, Q. Dumoulin, S. Sabau, Olivier Roche, Thomas Goetzmann |
---|---|
Rok vydání: | 2018 |
Předmět: |
030222 orthopedics
medicine.medical_specialty business.industry Bone stock Dentistry Retrospective cohort study Osseointegration Surgery 03 medical and health sciences 0302 clinical medicine Implant replacement Radiological weapon Orthopedic surgery medicine Orthopedics and Sports Medicine In patient 030212 general & internal medicine Implant business |
Zdroj: | Orthopaedics & Traumatology: Surgery & Research. 104:317-323 |
ISSN: | 1877-0568 |
DOI: | 10.1016/j.otsr.2017.11.024 |
Popis: | Introduction The PFMR® proximal femoral modular reconstruction implant (Protek, Sulzer Orthopedics, Switzerland) is a straight modular stem in sanded titanium with press-fit anchorage, intended to achieve spontaneous bone reconstruction following Wagner's principle. The aim of the present study was to analyze long-term clinical and radiological outcome. Material and method A single-center retrospective study included 48 PFMR stems implanted in 47 patients between 1998 and 2002. Results in this series were previously reported at 7 years’ follow-up. Clinical assessment used PMA and Harris scores. Radiologic assessment focused on stem stability and osseointegration, and bone stock following Le Beguec. Results Twenty-three patients were seen at a mean 14.5 years’ follow-up (13 deceased, 11 lost to follow-up), including 1 with bilateral implants, i.e., 24 stems. PMA and Harris scores, stem stability and osseointegration and bone stock were stable with respect to the 7-year findings. Radiology found 7 stem fractures in the Morse taper, i.e., in 29% of implants. Two of these cases required femoral implant replacement; 5 were asymptomatic. Discussion and conclusion Long-term outcome for PFMR stems was clinically and radiologically satisfactory for the 16 patients free of mechanical complications. The Morse taper fracture rate was high, and higher than reported elsewhere. The usual risk factors for implant fracture were not found in the present series. The modular design of the press-fit revision implant is its weak point; monoblock implants should be used in patients with good life-expectancy. Level of evidence IV (retrospective study). |
Databáze: | OpenAIRE |
Externí odkaz: |