Popis: |
In order to overcome the stop marketing by Biorad company of automated high performance liquid chromatograph with UV detection (Remedi), we developed a gas chromatography-mass spectrometry (GC-MS) to detect and to give an approximation of the overdose of molecules frequently encountered in drug intoxications. Therefore two hundred eighty seventeen blood samples were collected over a period of one year and allowed us to evaluate and compare the performance of these two techniques. As identification, GC-MS does not identify all molecules detected by Remedi in 24.2% of cases; there is a lack of sensitivity for opiates and the systematic absence of certain molecules such as betablockers. However, in 75.8% of cases the GC-MS detects all molecules found by Remedi and other molecules such as meprobamate, paracetamol, benzodiazepines and phenobarbital. The concentrations obtained are interpreted in terms of overdose showed 15.7% of discrepancy and 84.3% of concordance between the two techniques. The GC-MS technique described here is robust, fast and relatively simple to implement; the identification is facilitated by macro commands and the semi quantification remains manual. Despite a sequence of cleaning the column after each sample, carryover of a sample to the next remains possible. This technique can be used for toxicologic screening in acute intoxications. Nevertheless it must be supplemented by a HPLC with UV detection if molecules such as betablockers are suspected. |