Effects of B 0 eddy currents on imaging isocenter shifts in 0.35‐T MRI‐guided radiotherapy (MR‐IGRT) system

Autor: Benjamin C. Lewis, Taeho Kim, A. Curcuru, H. Michael Gach, Deshan Yang
Rok vydání: 2021
Předmět:
Zdroj: Medical Physics. 48:2929-2938
ISSN: 2473-4209
0094-2405
Popis: PURPOSE The purpose of this study was to measure gantry angle-related eddy currents in a 0.35-T MRI-Linac and determine if B0 (zeroth order) eddy currents are the primary cause of gantry angle-dependent imaging isocenter shifts vs other potential causes like B0 inhomogeneities and gradient (first order) eddy currents. For conventional Cartesian acquisitions, B0 eddy currents can cause imaging isocenter shifts along both phase encode and readout directions. Gradient eddy currents can cause spatial distortion along both the phase encode and readout directions. Center frequency offsets can cause imaging isocenter shifts along the readout direction that vary with readout gradient polarity. METHODS MRI-related eddy currents and imaging isocenter shifts were measured on a 0.35-T MRI-Linac at gantry angles from 0° to 330° in increments of 30° . All measurements were made after gradient shimming and center frequency tuning at each planned gantry angle. Eddy current and field homogeneity measurements were conducted using a 24-cm diameter spherical phantom. Gradient and B0 eddy currents were calculated from the free induction decays (FIDs) resulting from selective excitation of slices located ±5 cm from isocenter. B0 eddy currents were also calculated from FIDs acquired with nonselective excitation and compared with B0 eddy current values derived using selective excitation. B0 inhomogeneities and center frequency offsets were measured by acquiring FIDs with nonselective excitation. Imaging isocenter shifts were measured using a 33x33x10.5 cm3 uniformity linearity (grid) phantom and a 3D true fast imaging with steady-state precession (TrueFISP) sequence used in MRI-guided radiation therapy. Eddy currents were compared to vendor specifications and correlated with the imaging isocenter shifts. Measurements were conducted before and after the MRI-Linac's waveguide was replaced with an updated design to reduce eddy currents. RESULTS B0 eddy currents were highly correlated (r = 0.986, P <
Databáze: OpenAIRE