An inherently dangerous fluid warmer design
Autor: | Albert A Kalustian, Keegan L McDonnell, R. Mauricio Gonzalez, Karolina Brook, Mark J Facchin, Donald H. Lambert |
---|---|
Rok vydání: | 2021 |
Předmět: | |
Zdroj: | Journal of Clinical Monitoring and Computing. 36:909-915 |
ISSN: | 1573-2614 1387-1307 |
DOI: | 10.1007/s10877-021-00786-x |
Popis: | A Hotline® fluid warmer is a device commonly used by anesthesia providers in the operating room to warm and infuse blood products and large fluid volumes. The purpose of the fluid warmer is to counter heat loss, which occurs under anesthesia. Despite normal checks performed prior to its use, we discovered a breach in the fluid warming set attached to the Hotline® fluid warmer during blood administration. The breach contaminated the patient's sterile intravenous line. We describe the quality and safety processes we undertook in detail. We discuss the notion that monitoring alarms are an important safety feature of most modern devices utilized by anesthesia providers. We believe the Hotline® fluid warmer lacks a crucial monitor for detecting a breach within the fluid warming set, and therefore recommend the addition of an alarm to improve this device's safety. |