Vergleich verschiedener Atemgas-Anfeuchtersysteme für die Respiratortherapie Frühgeborener im Inkubator
Autor: | C. Roll, L. Hanssler, W. Tennhoff |
---|---|
Rok vydání: | 1994 |
Předmět: |
Artificial ventilation
Clinical tests medicine.medical_specialty Materials science medicine.medical_treatment Air humidity Humidity Incubator General Medicine Critical Care and Intensive Care Medicine Surgery Setpoint Anesthesiology and Pain Medicine Anesthesia Emergency Medicine medicine Respiratory system Endotracheal tube |
Zdroj: | AINS - Anästhesiologie · Intensivmedizin · Notfallmedizin · Schmerztherapie. 29:102-107 |
ISSN: | 1439-1074 0939-2661 |
DOI: | 10.1055/s-2007-996695 |
Popis: | Objective 7 humidifier systems for neonatology were subjected to a performance test in our laboratory. Design In 18 measurement combinations, flow (2,6,16 l/minute), respiratory gas temperature (34 and 37 degrees C) and incubator temperature (33, 35, 37 degrees C) were varied. Results In only 70.6% of 126 measurements the minimum international standard for the absolute humidity in the respiratory gas (30 mg H2O/l air) was achieved or exceeded. In 3 systems this level was not reached in 9 resp. 10 of 18 measurements. Only 1 device that functions according to the "membrane humidification" principle fulfills the requirements with regard to constant high air humidity. In 36 of 126 ( = 28.6%) measurement combinations the respiratory gas temperature near the endotracheal tube varied > 1 degrees C from the system temperature setpoint. Conclusion No controlled clinical tests regarding the importance and long term effects of different levels in the inspiratory air are so far available for the ventilatory treatment of neonates. |
Databáze: | OpenAIRE |
Externí odkaz: |