Electrochemical behaviour of persulphate on carbon paste electrodes modified with Prussian blue and analogous compounds

Autor: Kurt Kalcher, W. Ng, M. Weißenbacher, Herbert Greschonig, Wing Hong Chan, Anastasios Voulgaropoulos
Rok vydání: 1992
Předmět:
Zdroj: Fresenius' Journal of Analytical Chemistry. 344:87-92
ISSN: 1618-2650
0937-0633
Popis: Peroxodisulphate (“persulphate”) may be determined by direct current voltammetry (DCV) using carbon paste electrodes chemically modified with Prussian Blue, iron(III)-hexacyanoruthenate(II) or iron(III)-hexacyano-osmate(II). The determination is based on the exploitation of catalytic currents from the reduction of the modifiers. Best results are obtained for iron(III)-hexacyanoosmate(II) yielding a detection limit of 1 μg O22−/ml (as persulphate) when using HCl (10−2 mol/l) as supporting electrolyte and measuring in the DCV-mode. Electrodes doubly-modified with the osmium compound and a liquid anion-exchanger (Amberlite LA2) allow determinations of persulphate after preconcentration under open circuit conditions, in the presence of hydrogen peroxide and perborate, with a detection limit of 20 ng O22−/ml (as persulphate).
Databáze: OpenAIRE